Hydrogel Based Implant Resonator for Wireless pH Sensing
2025 IEEE International Symposium on Antennas and Propagation and North American Radio Science Meeting, AP-S/CNC-USNC-URSI 2025, Ottawa, Kanada, 13 - 18 Temmuz 2025, ss.1422-1425, (Tam Metin Bildiri)
- Yayın Türü: Bildiri / Tam Metin Bildiri
- Doi Numarası: 10.1109/ap-s/cnc-usnc-ursi55537.2025.11266562
- Basıldığı Şehir: Ottawa
- Basıldığı Ülke: Kanada
- Sayfa Sayıları: ss.1422-1425
- Boğaziçi Üniversitesi Adresli: Evet
Özet
This study introduces a pH-sensitive hydrogel-based implant resonator for real time, in-body sensing. The hydrogel's pH dependent dielectric and conductive properties enable its use as a passive microwave resonator, wirelessly interrogated via a wideband on-body reader antenna. The system, designed for chipless RFID-based communication, eliminates the need for active power sources in-body. It demonstrates high sensitivity to physiological pH changes, particularly relevant for postoperative infection detection, where wound pH shifts from 4.5-6 to 6.5-8 for healthy and infected respectively. Experimental results confirm that the dielectric properties of the hydrogel change with pH consistently with no hysteresis behavior, underscoring its potential for next-generation biomedical sensors in continuous health monitoring. It has been shown that the p H of the immediate surrounding of the hydrogel based sensor can be tracked with minimum of 5 MHz shifts for every 2-units of pH change.