A 24-GHz Doppler sensor system for cardiorespiratory monitoring
42nd Conference of the Industrial Electronics Society, IECON 2016, Florence, İtalya, 24 - 27 Ekim 2016, ss.5161-5164, (Tam Metin Bildiri)
- Yayın Türü: Bildiri / Tam Metin Bildiri
- Doi Numarası: 10.1109/iecon.2016.7793225
- Basıldığı Şehir: Florence
- Basıldığı Ülke: İtalya
- Sayfa Sayıları: ss.5161-5164
- Anahtar Kelimeler: 24 GHz, Antenna, Breathing sensor, Electronic hardware, K-band, Microwave, Radar, Remote life monitoring
- Boğaziçi Üniversitesi Adresli: Evet
Özet
In this paper, a 24-GHz Microwave band noninvasive sensor system for detecting and monitoring cardiorespiratory signals of patients is presented. The system consists of a sensor unit and a monitoring server. To demonstrate the proposal, a Doppler sensor unit operating at 24 GHz ISM band is fabricated. The fabricated radar unit shows below 6.2 % error for respiration rate when the distance between the patient and the radar system is below 1 m. The error increases to 7.3 % when the distance is increased to 2 m. To demonstrate the proposed medical sensor system, the life conditions of a subject is monitored. When an abnormality such as a stop breathing occurs, a warning signal is successfully generated automatically. In the case of an emergency, the health care workers can help the patients quicker.