An in-Body Far Field Wireless Power Transfer Link Under Homogenous Tissue Assumption


Salimitorkamani M., ÖZCAN B. F., Bilir A., Cakicioglu M., DUMANLI OKTAR S.

20th European Conference on Antennas and Propagation, EuCAP 2026, Dublin, İrlanda, 19 - 24 Nisan 2026, (Tam Metin Bildiri)

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Doi Numarası: 10.23919/eucap68105.2026.11612565
  • Basıldığı Şehir: Dublin
  • Basıldığı Ülke: İrlanda
  • Boğaziçi Üniversitesi Adresli: Evet

Özet

This paper presents the design, simulation, and experimental validation of an in-body far-field wireless power transfer link. The link consists an electrically small, low-profile implanted Huygens Source Antenna (HSA) resonating at 1.8 GHz and a wide-band on-body slot antenna that operates between 0.7 GHz and 2 GHz. A Capacitively Loaded Loop (CLL), driven with an SMA port via an unbalanced microstrip feed network acts as the primary source of the Magnetic Dipole (H Moment). The bottom layer hosts a pair of symmetric Near-Field Resonant Parasitic elements that are coupled to the central CLL, generating the orthogonal Electric Dipole (E Moment) component. The final dimensions of the antenna are 0.072 λ0 × 0.090 λ0 × 0.015 λ0. The HSA exhibits an efficiency of 0.18% confirming its suitability for power-efficient implant-tosurface communication and future integration into WPT systems. A channel transmission coefficient of -20.9 dB was achieved for an implant depth of 20 mm through phantom measurements.