A Novel Area Efficient Inductorless Super-Regenerative Receiver Front-End for Medical Brain Implants


Pekcokguler N., DÜNDAR G., Dehollain C.

19th International Conference on Synthesis, Modeling, Analysis and Simulation Methods, and Applications to Circuit Design, SMACD 2023, Funchal, Portekiz, 3 - 05 Temmuz 2023, (Tam Metin Bildiri) identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Doi Numarası: 10.1109/smacd58065.2023.10192188
  • Basıldığı Şehir: Funchal
  • Basıldığı Ülke: Portekiz
  • Anahtar Kelimeler: Epilepsy, Implant, Inductorless, Medical, Receiver, Super Regenerative
  • Boğaziçi Üniversitesi Adresli: Evet

Özet

A novel area and energy efficient inductorless super-regenerative receiver (SRR) is proposed for medical brain implants dedicated to epilepsy monitoring and treatment. The proposed SRR was fabricated in UMC CMOS 180nm technology. The SRR occupies only 0.072mm2 which corresponds to an order of magnitude area reduction compared to the state-of-the-art. The SRR achieves -70dBm sensitivity and the highest data rate is 4 Mbps. 3.22 nJ/b energy efficiency is obtained with 12.9mW power consumption from a 1.8V power supply.