Wireless Tracking of Cranial Bone Regeneration via Biodegradable Implant Degradation Monitoring


Salimitorkamani M., ÖZCAN B. F., DUMANLI OKTAR S.

20th European Conference on Antennas and Propagation, EuCAP 2026, Dublin, İrlanda, 19 - 24 Nisan 2026, (Tam Metin Bildiri)

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Doi Numarası: 10.23919/eucap68105.2026.11612072
  • Basıldığı Şehir: Dublin
  • Basıldığı Ülke: İrlanda
  • Boğaziçi Üniversitesi Adresli: Evet

Özet

Advances in additive manufacturing and biomaterials are ushering in a new era of cranial implants, where patientspecific 3D-printed biodegradable devices are becoming standard practice. Here, a partially conformal slot antenna is proposed to monitor the regeneration process of cranial bone with a biodegradable cranial implant. The on-body antenna operates at 2.4 GHz and is constructed using a substrate of Rogers 3010 and a flexible superstrate of graphite doped silicone. A biodegradable lattice-shaped implant with a radius of 12 mm is embedded within the cranial region and covered with a liquid layer to emulate realistic post-surgical conditions. By varying the implant radius from 12 mm to 2 mm, the bone regeneration and implant degradation is simulated. The resonant frequency of the on-body antenna changes as the defect heals due to the change in effective permittivity of the implanted region. It has been shown that a 2 mm change in radius results in at least 53 MHz change. This shift confirms the antenna's high sensitivity to bone regeneration progression and its potential for real-time, non-invasive monitoring of cranial healing.