Wireless Tracking of Cranial Bone Regeneration via Biodegradable Implant Degradation Monitoring
20th European Conference on Antennas and Propagation, EuCAP 2026, Dublin, İrlanda, 19 - 24 Nisan 2026, (Tam Metin Bildiri)
- Yayın Türü: Bildiri / Tam Metin Bildiri
- Doi Numarası: 10.23919/eucap68105.2026.11612072
- Basıldığı Şehir: Dublin
- Basıldığı Ülke: İrlanda
- Boğaziçi Üniversitesi Adresli: Evet
Özet
Advances in additive manufacturing and biomaterials are ushering in a new era of cranial implants, where patientspecific 3D-printed biodegradable devices are becoming standard practice. Here, a partially conformal slot antenna is proposed to monitor the regeneration process of cranial bone with a biodegradable cranial implant. The on-body antenna operates at 2.4 GHz and is constructed using a substrate of Rogers 3010 and a flexible superstrate of graphite doped silicone. A biodegradable lattice-shaped implant with a radius of 12 mm is embedded within the cranial region and covered with a liquid layer to emulate realistic post-surgical conditions. By varying the implant radius from 12 mm to 2 mm, the bone regeneration and implant degradation is simulated. The resonant frequency of the on-body antenna changes as the defect heals due to the change in effective permittivity of the implanted region. It has been shown that a 2 mm change in radius results in at least 53 MHz change. This shift confirms the antenna's high sensitivity to bone regeneration progression and its potential for real-time, non-invasive monitoring of cranial healing.